Biologische Veiligheidskast Klasse II A2 YR0090C-SS
Personeel · product · omgeving · drievoudige bescherming met HEPA H14

De Kalstein YR0090C-SS is een Klasse II Type A2 biologische veiligheidskast voor microbiologisch en biomedisch werk met BSL-1 tot BSL-3 agentia. HEPA H14 filters houden 99,995 % van deeltjes ≥ 0,3 µm tegen; 70 % van de lucht wordt gerecirculeerd en 30 % afgevoerd.
Interactive tour · Explore the YR0090C-SS
Every component, engineered for containment
Six precision-engineered systems working together — press play for a guided tour or click any point on the cabinet to inspect it in detail.
Inspecteerbare onderdelen
- Dual HEPA filtrationAir filtration
Supply and exhaust HEPA filtration retains particles ≥ 0.3 µm for Class II A2 containment.
- Digital control panelControl
Fan, UV lamp and LED lighting are operated from one digital interface.
- UV disinfection lampDecontamination
Germicidal UV supports chamber decontamination between operations.
- LED work lightingLighting
Waterproof LED lamps provide uniform visibility across the work zone.
- Tempered glass sashSafety
The front sash works as a physical barrier between operator and samples.
- AISI 304 work zoneMaterials
Stainless work surface designed for cleaning and decontamination.
Belangrijkste kenmerken
- Klasse II Type A2 gecertificeerd voor BSL-1 tot BSL-3
- HEPA H14 downflow- en afvoerfilters — 99,995 % bij 0,3 µm
- Downflow 0,33 m/s en inflow 0,55 m/s voor stabiele laminaire insluiting
- RVS AISI 304 werkkamer met afgeronde hoeken
- Gemotoriseerd frontraam met veiligheidsalarm
- Microprocessor met realtime luchtstroommonitoring
Drievoudige bescherming · engineering-lagen
Personnel protection
Inflow curtain at 0.55 m/s prevents aerosols from escaping the work zone, isolating the operator from BSL-1 to BSL-3 agents.
- Inflow barrier
- Front sash interlock
- Audible alarms
- Inflow velocity
- 0.55 m/s
- Face velocity
- ≥ 100 fpm
- Max biosafety level
- BSL-3
Product protection
HEPA-filtered downflow at 0.33 m/s bathes the sample in ISO 5 laminar air, keeping the culture free from environmental contamination.
- Downflow laminar
- ISO 5 work zone
- HEPA H14 downflow
- Downflow velocity
- 0.33 m/s
- Work zone class
- ISO 5
- Retention @ 0.3 µm
- 99.995%
Environmental protection
30% of the air is exhausted through a second HEPA H14 filter, releasing only clean air back into the laboratory or ducted exhaust.
- HEPA H14 exhaust
- Top duct connection
- Room air quality preserved
- Exhaust ratio
- 30%
- Exhaust filtration
- HEPA H14
- Noise level
- ≤ 62 dB
Luchtstroommodi · wat gebeurt in elke richting
- Downflow — HEPA downflow at 0.33 m/s
Vertical laminar air descends from the supply HEPA over the entire work zone, keeping the sample under ISO 5 conditions.
- Inflow — Inflow curtain at 0.55 m/s
Room air enters through the front grille at 0.55 m/s, creating an air barrier that stops aerosols from escaping.
- Exhaust — HEPA exhaust through the top duct
30% of the internal air is HEPA H14 filtered and released through the top duct.
- Full cycle — Simultaneous triple protection
Downflow, inflow and exhaust combine to protect personnel, product and environment at the same time.
Technische specificaties
| Model | YR0090C-SS |
|---|---|
| Type | Class II Type A2 |
| Filtratie | HEPA H14 (99.995% @ 0.3 µm) |
| Neerwaartse snelheid | 0.33 m/s |
| Inlaatsnelheid | 0.55 m/s |
| Afvoerverdeling | 30% exhaust · 70% recirculation |
| Buitenafmetingen | 1350 × 800 × 2250 mm |
| Werkkamer | 1130 × 590 × 660 mm |
| Geluidsniveau | ≤ 62 dB |
| Verlichting | ≥ 1000 lux |
| Voeding | 220 V / 50-60 Hz · 550 W |
Toepassingen
- Microbiologie en celkweek
- Virologie en BSL-3 onderzoek
- Klinische diagnostiek en IVF
- Farmaceutische bereiding
- Moleculaire biologie en PCR
Praktijkscenario’s · begeleide flows
Cell culture and tissue work
Sterile cell culture in a Class II Type A2 environment for cell lines, primary cultures and tissue processing.
- Sanitize the work surface with 70% ethanol
- Start the cabinet and wait for stable airflow
- Transfer materials through the front opening
- Work in the front-to-back laminar zone
- Decontaminate before shutdown
Molecular biology and PCR setup
Prepare PCR reactions and molecular assays free from cross-contamination.
- UV decontaminate for 15 minutes
- Load pipettes, tips and reagents
- Prepare master mix in the clean zone
- Aliquot into reaction tubes
- Clean and log the session
BSL-3 pathogen handling
Manipulation of high-risk agents when integrated in a BSL-3 laboratory suite.
- Verify certification and airflow reports
- Don PPE and follow SOPs
- Introduce material via pass-through
- Perform the procedure inside the containment
- Autoclave waste and decontaminate the chamber
Ideal for applications in
Where the YR0090C-SS performs
Explore each field the YR0090C-SS was engineered for — click any industry to reveal the specific benefits, key metrics and how the cabinet supports the workflow.
Microbiology laboratories
Culture handling, colony isolation and pathogen manipulation under sterile conditions.
key benefits
- Sterile culture handling
- Colony isolation & sub-culturing
- Pathogen manipulation up to BSL-2
- Cleanliness
- ISO 5
- Downflow
- 0.45 m/s
- Continuous
- 24/7
Biotechnology & research
Cell culture, molecular biology and genetic engineering with sample-to-operator isolation.
key benefits
- Primary & stable cell lines
- Transfection / transduction
- CRISPR & genetic engineering
- LED lighting
- 2×10 W
- Stainless
- AISI 304
- Filtration
- Dual HEPA
Pharmaceutical industry
Sterile compounding, formulation R&D and QC of aseptic preparations.
key benefits
- USP 797 aseptic compounding
- Chemotherapy admixtures
- Small-batch sterile filling
- Tipo A2
- Class II
- Recirc / exhaust
- 70/30
- Stainless
- AISI 304
Clinical & diagnostic labs
Processing of blood, tissue and clinical specimens with biological containment.
key benefits
- Blood & serum processing
- Tissue & biopsy handling
- Molecular diagnostics prep
- Containment
- BSL-2
- Decontamination
- UV lamp
- @ 0.3 µm
- ≥99.999%
Biohazardous materials
Safe manipulation of BSL-2 agents, low-volatility chemicals and infectious samples.
key benefits
- BSL-2 infectious agents
- Trace volatile chemicals
- Radionuclide trace work
- HEPA exhaust
- 30%
- Tempered glass
- 5 mm
- Plenum
- Sealed
Fast quotation · 15 seconds
Bring the YR0090C-SS to your laboratory
Receive a full quotation with installation and validation, delivered by Kalstein specialists.
Veelgestelde vragen
Welk biosafety-niveau?
Gecertificeerd voor BSL-1, BSL-2 en BSL-3 bij installatie/validatie volgens EN 12469 / NSF 49.
Klasse II Type A2?
Ja. 70 % HEPA-recirculatie, 30 % HEPA-afvoer via bovenkanaal.
Filtratie?
HEPA H14 op downflow en afvoer, retentie ≥ 99,995 % bij 0,3 µm.
On-site validatie?
Ja, volgens EN 12469 en NSF/ANSI 49.