Description
The Small Sample Viscometer YR05865 is a state-of-the-art digital rotary viscometer designed for precision and efficiency. Utilizing a coaxial cylinder structure, it requires only a minimal sample size of 7ml to 13ml per measurement, except for rotor 0 which needs 21ml. This makes it ideal for valuable and rare sample measurements. It features an optional constant temperature water jacket, ensuring excellent temperature control with a constant temperature tank. The viscometer is user-friendly, offering accurate and reliable measurements through its digital display.
Market Price
Viscometers like the YR05865 typically range in price depending on various features and brand reputation. For this model, the market price falls between approximately $2300 and $3100, offering great value for its advanced functionalities. It is positioned competitively in the market, reflecting its high accuracy and performance standards.
Frequently Asked Questions
What sample size does the YR05865 require? The viscometer uses as little as 7ml for testing, ensuring minimal wastage of precious samples.
Can the device be connected to other instruments? Yes, it features two RS232 interfaces that allow direct connections to a micro printer and a computer for seamless data transfer.
Advantages and Disadvantages
The YR05865 offers several advantages, including high precision and low sample requirement, making it perfect for precious liquid analysis. Its user-friendly interface and digital readouts simplify operations. However, the complexity of its advanced features may require additional training for optimal use.
Product Use in the Field
The Small Sample Viscometer YR05865 is ideal in laboratory settings where precise viscosity measurement of small batches is crucial, such as in pharmaceuticals and specialty chemicals. Its design accommodates rare and valuable samples, perfect for research and quality control.
Recommendations
To maximize the effectiveness of the YR05865, ensure routine calibration and cleanliness of the instrument to preserve measurement accuracy. Regular training sessions on its operation can further enhance user proficiency and measurement precision.
Features
- Timing measurement function
- Coaxial cylinder structure
- Using a spindle rotor, the sample size required for testing is very small
- Two RS232 interfaces with direct connection to micro printer and computer
Technical Specifications
| Model | YR05865 |
| Display mode | Digital Display |
| Speed (r/min) | 0.3/0.6/1.5/3/6/10/12/15/20/25/30/40/50/60/70/80/90/100 Total 18 |
| Rotor | 18#,25#,31#,34# (standard configuration) |
| Measuring Range | 3-1.6M (mPa﹒s) |
| Sample Size | 18# rotor:3 –10K; 7ml 25# rotor:480 –1.6M; 9ml 31# rotor:30 –100K; 10.5ml 34# rotor:60 –200K; 11ml |
| Measurement Error | ±3% (Newtonian liquid) |
| Repetitive Error | ±1.5% (Newtonian liquid) |
| Output Interface | Two RS232 interfaces to connect printer and computer |
| Temperature Probe Interface | Standard Configuration |
| Outline Dimension | 300 × 300 × 450(mm) |